IOTA LSR 3950 Series (50–70 Shore A) is a medical-grade two-component addition-cure liquid silicone rubber specially designed for laryngeal masks. Main features include excellent biocompatibility, two-component 1:1 mixing ratio, excellent mold release, fast heat curing, and excellent mechanical properties. The application direction is laryngeal masks and other medical devices. Technical data give three hardness grades: 3950-50, 3950-60, and 3950-70, with cured hardness of 50, 60, and 70; tensile strength of about 10, 9, and 9 MPa; elongation of about 500%, 400%, and 340%; tear strength of about 40, 35, and 32 kN/m; density of about 1.12, 1.12, and 1.14 g/cm³; and a note stating compliance with ISO 10993.
The curing time section only says “see curing time” and does not give a specific temperature-time table; actual curing conditions should be obtained from the supplier or confirmed with the equipment and product structure.
Selection should not rely only on “medical grade,” “ISO 10993 compliance,” or “hardness 50–70.” First confirm the specific medical device and contact mode, sterilization method and cycles, product structure and thickest section, mold and release requirements, same-batch A/B 1:1 metering and mixing, whether the material is thoroughly stirred, curing temperature and time, hardness grade selection, and the biocompatibility test items and regulatory compliance documents required by the project. This article does not equate “ISO 10993 compliance” with final medical device compliance, nor does it extend “for laryngeal masks” to all medical devices. IOTA can provide liquid silicone rubber raw materials and corresponding product data; the company does not provide complete formulations or formulation development services.
The series is used for laryngeal masks and other medical devices. Different medical devices focus on different points: laryngeal masks on softness, sealing, biocompatibility, and sterilization resistance; other medical devices on hardness, rebound, tear strength, and dimensional stability. Within the series, hardness ranges from 50 to 70 across 3950-50 to 3950-70. Selection should start with device structure, contact mode, sterilization method, and assembly requirements, not just “medical grade” or “pick a hardness.”
| What to confirm | What to specify in an inquiry or sample request |
|---|---|
| Medical application and contact mode | Laryngeal mask or other medical device; body contact site, contact duration, short-term or long-term contact |
| Regulation and biocompatibility | Specific ISO 10993 test items, regulatory requirements, customer restricted-substance list; test reports and compliance documents required |
| Sterilization method | Sterilization method (e.g., ethylene oxide, irradiation, steam), number of cycles, post-sterilization performance requirements |
| Device structure and mold | Device dimensions, thickest section, mold material, release method, whether release agent is used |
| Mixing and metering | Two-component A:B = 1:1; only same-batch A/B may be mixed; metering accuracy, mixing uniformity, and injection equipment |
| Material pretreatment | Whether thoroughly stirred before transfer to container; whether fillers or colorants are uniformly dispersed |
| Curing conditions | Curing temperature and time; the data only say “see curing time,” so obtain from the supplier and confirm with product thickness and equipment |
| Hardness grade selection | 3950-50/60/70; select with device structure, feel, sealing, assembly force, and post-sterilization performance |
| Acceptance criteria | Hardness, tensile strength, tear strength, elongation, compression set, rebound, sterilization resistance, biocompatibility, etc. |
| Sample preparation | Sample size, thickness, mold condition, vulcanization conditions, secondary heating, whether evaluated after sterilization, sample quantity |
If the customer already has a test specification, obtain the original document first. When criteria are missing, confirm them with the project owner and test laboratory before making samples, so that both sides do not discover different interpretations of “pass” after testing.
IOTA LSR 3950 Series uses a two-component 1:1 mixing ratio. Important notes: the series is platinum-catalyzed, with the platinum catalyst in Component A; strict care is required during use to avoid catalyst poisoning. Only Component A and Component B of the same batch number may be mixed together. To ensure optimum flow and uniformity, the material must be thoroughly stirred before transfer into a container.
The curing time section only says “see curing time” and does not provide a specific temperature-time table. Therefore, sample-request and acceptance documents should not simply apply conditions from other series. Obtain the current batch curing recommendation from the supplier and confirm it with the thickest section, mold temperature, injection amount, and equipment efficiency. A tack-free or briefly set film does not necessarily mean the final curing condition has been reached.
The product note states compliance with ISO 10993. ISO 10993 is a set of biocompatibility evaluation standards containing multiple specific test items; raw-material “ISO 10993 compliance” should be understood as results under specific test items and is not the same as automatic compliance of the final medical device. For laryngeal masks and other medical devices, obtain the corresponding test reports and regulatory compliance documents, and evaluate according to the specific contact site, contact duration, sterilization method, and customer requirements. The final device must also be evaluated with formulation, process, additives, vulcanization conditions, and sterilization conditions.
“For laryngeal masks” is a product positioning description and should not be extended to mean suitability for all medical devices. Other medical devices should separately confirm hardness, biocompatibility, sterilization resistance, and regulatory requirements.
Technical data: 3950-50/60/70 cured hardness 50/60/70, tensile strength 10/9/9 MPa, elongation 500%/400%/340%, tear strength 40/35/32 kN/m, and density about 1.12/1.12/1.14 g/cm³. As hardness increases, elongation decreases and tear strength slightly decreases. Selection should be confirmed with the laryngeal mask’s softness, sealing, assembly force, and post-sterilization performance, not hardness alone.
Mold release and mechanical properties should also be evaluated with the specific mold and process. Release performance depends on mold material, surface condition, device structure, and cure degree. Whether mechanical properties are retained after sterilization should be tested separately.
Obtain the customer’s service conditions, regulatory requirements, and acceptance criteria; confirm whether the medical application, contact mode, sterilization method, ISO 10993 items, and hardness grade requirements are targets or existing results.
Specify device structure and thickest section, mold, same-batch A:B = 1:1 mixing/metering, material stirring, curing temperature and time, sterilization arrangement, and hardness grade.
Check the current IOTA LSR 3950 Series data, batch data, and biocompatibility reports, then have the user make representative samples and keep records of mixing, injection, vulcanization, release, sterilization, and post-treatment.
Arrange tests as agreed; keep sample identity, test conditions, observation photos, and evaluation results in the report. Before expanding use, confirm whether the samples represent actual devices.
IOTA/Anhui Iota Silicone Oil Co., Ltd. positions itself as a solution provider across the organosilicon industry chain and can provide liquid silicone rubber raw materials and corresponding product data. When communicating, provide the medical device, contact mode, sterilization method, regulatory requirements, mold, and curing equipment together to facilitate candidate screening; the company does not provide complete formulations or formulation development services.
No. ISO 10993 contains multiple specific test items. Raw-material compliance with specific items is not the same as automatic compliance of the final medical device; evaluation must also consider contact site, contact duration, sterilization method, and regulatory requirements.
Not recommended. Obtain the current batch curing recommendation from the supplier and confirm it with the thickest section, mold temperature, injection amount, and equipment efficiency.
Select with the laryngeal mask’s softness, sealing, assembly force, device structure, and post-sterilization performance, not hardness alone. As hardness increases, elongation decreases and tear strength slightly decreases.
The data explicitly require that only Component A and Component B of the same batch number be mixed together. Mixing different batches may affect metering, curing, and property consistency; follow the customer’s raw-material management and supplier requirements.
The platinum catalyst is in Component A. Strict care is required during use to avoid contact with substances known to poison platinum catalysts. Specific material compatibility should be included in the customer’s raw-material approval and site management requirements.
The data do not give post-sterilization property data. Hardness, tensile strength, tear strength, rebound, and dimensional changes should be tested separately with the actual sterilization method and number of cycles; unsterilized data cannot be used to infer post-sterilization performance directly.
No direct extrapolation. Other medical devices should separately confirm hardness, biocompatibility, sterilization resistance, and regulatory requirements, and obtain corresponding test evidence.