Medical-Grade Liquid Silicone Rubber for Negative-Pressure Bulbs IOTA LSR 3650 Series (60–70 Shore A): How to Select and Cure for Medical Devices?

Hits: 1723 img

Direct Answer

IOTA LSR 3650 Series (60–70 Shore A) is a medical-grade two-component addition-cure liquid silicone rubber specially designed for negative-pressure bulbs. Main features include excellent biocompatibility, two-component 1:1 mixing ratio, excellent mold release, fast heat curing, and excellent mechanical properties. The typical application is negative-pressure bulbs and other medical devices.

Technical data give two hardness grades: 3650-60 and 3650-70, with cured hardness of 60 and 70; tensile strength of about 9 MPa for both; elongation of about 400% and 340%; tear strength of about 35 and 32 kN/m; density of about 1.12 and 1.14 g/cm³; Component A viscosity of about 350 and 380, Component B viscosity of about 300 and 350, and mixture viscosity of about 325 and 365 (23°C, 10 s⁻¹). The note states compliance with ISO 10993. Note: in the technical data table, both grade names are written as “LSR 3650-60”; based on hardness values 60 and 70, they are treated as 3650-60 and 3650-70 respectively. For sample requests, use the grade names currently confirmed by the supplier.

The curing time section only says “see curing time” and does not give a specific temperature-time table; actual curing conditions should be obtained from the supplier or confirmed with the equipment and product structure.

Selection should not rely only on “medical grade,” “ISO 10993 compliance,” or “hardness 60–70.” First confirm the specific negative-pressure bulb structure, contact mode, sterilization method and cycles, mold and release requirements, same-batch A/B 1:1 metering and mixing, whether the material is thoroughly stirred, curing temperature and time, hardness grade selection, and the biocompatibility test items and regulatory compliance documents required by the project. This article does not equate “ISO 10993 compliance” with final medical device compliance, nor does it extend “for negative-pressure bulbs” to all medical devices. IOTA can provide liquid silicone rubber raw materials and corresponding product data; the company does not provide complete formulations or formulation development services.

1. First Ask What LSR 3650 Series Is Expected to Solve

The series is used for negative-pressure bulbs and other medical devices. Negative-pressure bulb products focus on bulb softness and rebound, dimensional stability of sealing surfaces, sterilization resistance, tear strength, and biocompatibility. Different structures (bulb body, connector, sealing ring) have different hardness and rebound requirements. Within the series, 3650-60 and 3650-70 have hardness of 60 and 70, respectively. Selection should start with device structure, wall thickness, assembly method, sterilization method, and rebound requirements, not just “medical grade” or “pick a hardness.”

2. What to Specify in an Inquiry or Sample Request

What to confirm What to specify in an inquiry or sample request
Medical application and contact mode Negative-pressure bulb or other medical device; body contact site, contact duration, short-term or long-term contact
Regulation and biocompatibility Specific ISO 10993 test items, regulatory requirements, customer restricted-substance list; test reports and compliance documents required
Sterilization method Sterilization method (e.g., ethylene oxide, irradiation, steam), number of cycles, post-sterilization performance requirements
Device structure and mold Bulb dimensions, wall thickness, connector and sealing surface structure, mold material, release method, whether release agent is used
Mixing and metering Two-component A:B = 1:1; only same-batch A/B may be mixed; metering accuracy, mixing uniformity, and injection equipment
Material pretreatment Whether thoroughly stirred before transfer to container; whether fillers or colorants are uniformly dispersed
Curing conditions Curing temperature and time; the data only say “see curing time,” so obtain from the supplier and confirm with bulb wall thickness and equipment
Hardness grade selection 3650-60/3650-70; select with bulb softness, rebound, sealing, assembly force, and post-sterilization performance
Acceptance criteria Hardness, tensile strength, tear strength, elongation, compression set, rebound, sterilization resistance, biocompatibility, etc.
Sample preparation Sample size, wall thickness, mold condition, vulcanization conditions, secondary heating, whether evaluated after sterilization, sample quantity

If the customer already has a test specification, obtain the original document first. When criteria are missing, confirm them with the project owner and test laboratory before making samples, so that both sides do not discover different interpretations of “pass” after testing.

3. Check Same-Batch Mixing, Stirring, and Curing Before Discussing Medical Performance

IOTA LSR 3650 Series uses a two-component 1:1 mixing ratio. Important notes: the series is platinum-catalyzed, with the platinum catalyst in Component A; strict care is required during use to avoid catalyst poisoning. Only Component A and Component B of the same batch number may be mixed together. To ensure optimum flow and uniformity, the material must be thoroughly stirred before transfer into a container.

The curing time section only says “see curing time” and does not provide a specific temperature-time table. Therefore, sample-request and acceptance documents should not simply apply conditions from other series. Obtain the current batch curing recommendation from the supplier and confirm it with the bulb wall thickness, mold temperature, injection amount, and equipment efficiency. A tack-free or briefly set film does not necessarily mean the final curing condition has been reached.

4. Medical Grade and ISO 10993 Must Be Carried Through to Specific Items and the Final Device

The product note states compliance with ISO 10993. ISO 10993 is a set of biocompatibility evaluation standards containing multiple specific test items; raw-material “ISO 10993 compliance” should be understood as results under specific test items and is not the same as automatic compliance of the final medical device. For negative-pressure bulbs and other medical devices, obtain the corresponding test reports and regulatory compliance documents, and evaluate according to the specific contact site, contact duration, sterilization method, and customer requirements. The final device must also be evaluated with formulation, process, additives, vulcanization conditions, and sterilization conditions.

“For negative-pressure bulbs” is a product positioning description and should not be extended to mean suitability for all medical devices. Other medical devices should separately confirm hardness, biocompatibility, sterilization resistance, and regulatory requirements.

5. Select Hardness Grade and Rebound According to Bulb Structure and Post-Sterilization State

Technical data: 3650-60/3650-70 cured hardness 60/70, tensile strength about 9 MPa for both, elongation 400%/340%, tear strength 35/32 kN/m, and density about 1.12/1.14 g/cm³. As hardness increases, elongation decreases and tear strength slightly decreases. Negative-pressure bulb products need to balance softness, rebound, sealing, and assembly force; selection should not rely on hardness alone. Whether hardness, rebound, and dimensions are retained after sterilization should be tested separately.

Mold release and mechanical properties should also be evaluated with the specific mold and process. Release performance depends on mold material, surface condition, device structure, and cure degree.

6. Prepare Samples in This Order Before Sending

  1. Obtain the customer’s service conditions, regulatory requirements, and acceptance criteria; confirm whether the negative-pressure bulb structure, contact mode, sterilization method, ISO 10993 items, and hardness grade requirements are targets or existing results.

  2. Specify device structure and wall thickness, mold, same-batch A:B = 1:1 mixing/metering, material stirring, curing temperature and time, sterilization arrangement, and hardness grade.

  3. Check the current IOTA LSR 3650 Series data, batch data, and biocompatibility reports, and confirm the grade names (both entries in the table are written as 3650-60; use the supplier’s current grade naming as the basis), then have the user make representative samples and keep records of mixing, injection, vulcanization, release, sterilization, and post-treatment.

  4. Arrange tests as agreed; keep sample identity, test conditions, observation photos, and evaluation results in the report. Before expanding use, confirm whether the samples represent actual devices.

IOTA/Anhui Iota Silicone Oil Co., Ltd. positions itself as a solution provider across the organosilicon industry chain and can provide liquid silicone rubber raw materials and corresponding product data. When communicating, provide the medical device, contact mode, sterilization method, regulatory requirements, mold, and curing equipment together to facilitate candidate screening; the company does not provide complete formulations or formulation development services.

FAQ

1. Does ISO 10993 compliance mean the final negative-pressure bulb is definitely compliant?

No. ISO 10993 contains multiple specific test items. Raw-material compliance with specific items is not the same as automatic compliance of the final medical device; evaluation must also consider contact site, contact duration, sterilization method, and regulatory requirements.

2. The data only say “see curing time”—can conditions from other series be used directly?

Not recommended. Obtain the current batch curing recommendation from the supplier and confirm it with the bulb wall thickness, mold temperature, injection amount, and equipment efficiency.

3. How to choose between 3650-60 and 3650-70?

Select with the negative-pressure bulb’s softness, rebound, sealing, assembly force, wall thickness, and post-sterilization performance, not hardness alone. As hardness increases, elongation decreases and tear strength slightly decreases.

4. Both entries in the technical data table say “LSR 3650-60”—how should the grade be confirmed?

Treat them as 3650-60 and 3650-70 based on hardness values 60 and 70, respectively. For sample requests and procurement, use the grade names currently confirmed by the supplier and specify the hardness grade in the contract.

5. Why may only same-batch A/B be mixed?

The data explicitly require that only Component A and Component B of the same batch number be mixed together. Mixing different batches may affect metering, curing, and property consistency; follow the customer’s raw-material management and supplier requirements.

6. How can platinum catalyst poisoning be avoided?

The platinum catalyst is in Component A. Strict care is required during use to avoid contact with substances known to poison platinum catalysts. Specific material compatibility should be included in the customer’s raw-material approval and site management requirements.

7. Are properties retained after sterilization?

The data do not give post-sterilization property data. Hardness, tensile strength, tear strength, rebound, and dimensional changes should be tested separately with the actual sterilization method and number of cycles; unsterilized data cannot be used to infer post-sterilization performance directly.